Loading...

Product Details

FLAME PHOTOMETER 128 x 64 GRAPHICAL LCD DISPLAY ( NU-341 )

With Sodium (Na) and Potassium (K) Filters

Technical Specification

      Mode               

General

Serum

Urine

Na : 0 to 100 ppm Up to 250 meq/L, 1:100 dil

Na : 0 to 200 ppm, 1 : 100 dil

Na : 0 to 250 ppm, 1 : 100 dil

K : 0 to 100 ppm Up to 250 meq/L, 1 : 100 dil

K : 0 to 10 ppm, 1 : 100 dil

Na : 0 to 100 ppm, 1 : 100 dil

Ca : 0 to 100 ppm Up to 250 meq/L, 1 : 100 dil

-

Ca : 0 to 100 ppm, 1 : 2 dil

Li : 0 to 100 ppm Up to 250 meq/L, 1 : 100 dil

Li : 0 to 2 ppm, 1 : 5 dil

-

      Sensitivity

General : Na : 0.5 ppm, K : 0.5 ppm, Li : 0.5 ppm, Ca : 15 ppm

      Filters

Na & K

Ca & Li ( Optional if Ordered )

      Resolution                         

0.1 ppm / meq

      Reproducibility

< 1% coefficient of variation for 20 consecutive samples using 10 ppm, 
Na set as maximum standard.

      Curve Fit Accuracy

< 2% error when 3 ppm Na/K and 5 ppm Li, set as max. Standard.

      Display

128 x 64 Graphical LCD

      Average Time

In built Software

      Flame System

LPG and Dry Oil free Air

      Detector

Highly Sensitive Silicon Photo Diode

      Calibration

Upto 5 point calibration with curve fitting software

      Gas Control

Adjustable with knob

      Flame Failure

Auto Detection

      Gas Cut Off

Gas Feeding, auto cut off in case of Power Failure

      Ignition System

Auto Ignition

      Printer

Provision for attachment of Dot Matrix Printer

      RS - 232

Provision for Data Transfer to PC

      Power

230 V ± 10%, 50 Hz AC

      Dimension

280 x 250 x 210 mm ( L x W x H ) approx.

      Weight

7.4 Kg approx.

 

Compressor Unit

      Air Supply                

By oil free mini Compressor unit with pressure Regulator

      Combustion

LPG controlled by precision Regulator

      Power

230 V ± 10%, 50 Hz AC

      Dimension

290 x 250 x 210 mm ( L x W x H ) approx.

      Weight

8 Kg approx.

 

Standard Accessories

      Na & K Filter

                                 1 Nos. each, Factory fitted                       

      Chimney

1 Nos.

      Air & Gas Tubing

1 Set

      Capillary Tubes

3 Nos.

      RS-232 Cable

1 Nos. ( Optional : If Printer Interface is required )

      Mains Lead

1 Nos.

      Operation Manual

1 Nos.

      There are some additional accessories, which are essential for the operating of Instrument and has to be arranged by Customer as listed below.
      1. LPG Gas Cylinder
      2. Gas Regulator, required to be attached at the top/nozzle of the Gas Cylinder.
      3. Gas Pipe for LPG connection, from Gas Cylinder/Regulator to the Instrument.

      Optional : 1. "Ca" & "Li" Filter can be supplied on extra cost of each.